Spin Echo sequence is based on repetition of 90° and 180° RF pulses. Spin Echo sequence have two parameters:
Each tissue has a specific proton density, T1 and T2 time. The NMR signal depends on these 3 factors.
After time T1, longitudinal magnetization has returned to 63 % of its final value. T1 defines the recovery rate of longitudinal magnetization.
For example, here are longitudinal magnetization recovery curves for 2 tissues A and B with different T1s.
After time T2, transverse magnetization has returned to 37 % of its initial value. T2 defines the decay rate of transverse magnetization.
For example, here are transverse magnetization decay curves for 2 tissues A and B with different T2s.
How do TR and TE modify tissue signals?
Let's consider 2 tissues A and B with different T1s. If TR is very long, even if tissue A has a longuer T1 than tissue B, the longitudinal magnetization of both tissues will recover completely before the next excitation.
Thus, the transverse magnetization amplitude will be the same for both tissues after each excitation.
If TR is short and if tissue A has a longer T1 than tissue B, the longitudinal magnetization
of tissue A will recover less than the longitudinal magnetization of tissue B.
Thus, the transverse magnetization amplitude of tissue B will be higher after the next excitation.
In the graphs below :
By setting the TR to short values, tissue contrast will depend on differences in longitudinal magnetization recovery (T1).
By setting the TR to long values, the T1 effect on tissue contrast will be reduced. If TE is long enough, differences in transverse relaxation will alter tissue contrast (the T2 effect).
(But if TE is too long, the signal will have disappeared !)
To sum up:
Nearly all MR image display tissue contrasts that depend on proton density, T1 and T2 simultaneously. PD, T1 and T2 weighting will vary with sequence parameters, and may differ between different tissues in the same image.
The following table shows T1 and T2 relaxation times for various tissues at 1.5 T.
For example:
| T1 (ms) | T2 (ms) | |
| Water | 3000 | 3000 |
| Gray matter | 810 | 100 |
| White matter | 680 | 90 |
| Liver | 420 | 45 |
| Fat | 240 | 85 |
| Gadolinium | Reduces T1 and T2 times | |
In clinical practice:
It is your turn now! Change the TR and TE sequence parameters and the T1 / T2 times of the tissues and observe the contrast and acquisition time variations.
Don't forget: a good MRI sequence gives high tissue contrast but lasts the shortest time possible!